Description
ECG90A is an Electrocardiograph which can collect 12-lead ECG signal simultaneously and print waveform by the thermal printing system. It features in, recording and displaying ECG waveform in AUTO/Manual mode, measuring and diagnosing ECG waveform parameters automatically, prompting for “Lead off” and “Lack of paper”, multi-language interface, case database management.
Note: The device sold to the countries and regions that have FDA certification requirement does not support automatic analysis function.
1)Sync collection for 12-lead ECG, adopt digital signal processing technology and get high-quality ECG waveform via power frequency filter, baseline filter and EMG filter of ECG signal.
2)Particular and high-accuracy digital filter eliminates baseline drift, no distortion to ECG waveform, greatly enhances anti-baseline drift capacity, convenient for waveform judgment.
3)Display of 1/3/6/12-lead ECG, print mode, sensitivity, paper speed and filter state, etc. on one screen, convenient for contrastively diagnosing.
4)Real-time and continuously record clear and accurate ECG waveform and annotation character(including lead mark, sensitivity, paper speed and filter state, etc.).
5)Adopt high-resolution thermal printing system, no need for any adjustments. Recording frequency: up to 150 Hz.
6)With the functions of auto-analysis and auto-diagnosis for routine ECG parameters, provide measurement results and auto-diagnosis conclusion for HR, P-R interval, P Duration, QRS Duration, Q-T interval, Q-Tc, P Axis, QRS Axis, T Axis, R(V5), S(V1), R(V5)+S(V1), etc. which reduces the doctor’s burden.
7)Optional SD card stores up to 100 cases, which is convenient for doctor to review case and statistic.
8)In review mode, case waveform stored can be checked, measurement parameter and diagnosis information can be automatically analyzed.
9)Multi- language(Chinese, English) interface and report.
10)Optional ECG-Sync software, connect with PC by USB2.0 interface to form a ECG workstation, import real-time data to the software for analyzing and report printing, and old cases stored in Electrocardiograph can be uploaded to PC.
11)Power supply: AC/DC and built-in rechargeable lithium battery. In optimal DC state, up to 4-hour standby time, continuous print 90-minute and 150-ECG, which meets the requirements of visiting a patient at home and body examination.
Input mode: floating and defibrillation protection
Frequency response:
Rated input amplitude Input frequency and waveform Relative output response
1.0 0.67Hz~40Hz, Sine wave ±10%a
0.5 40Hz~100Hz, Sine wave +10 %, -30 %a
0.25 100Hz~150Hz, Sine wave +10 %, -30 %a
0.5 150 Hz ~ 500 Hz, Sine wave +10 %, -100 %a
1.5 ≤1Hz,200ms, Triangle wave +0 %, -10 %b
a relative to 10Hz b relative to 200 ms
Time constant:≥ 3.2s
CMRR: >105dB
Filter: power frequency(AC50/60 Hz), EMG(25 Hz/35 Hz (-3 dB)), baseline drift filter
Recording way: thermal printing system
Specification of recording paper: 50 mm(W)×20 m(L) high-speed thermal paper
Time base selection(paper speed): 6.25 mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s, error: ±5%
Gain control(sensitivity): 2.5, 5, 10, 20 mm/mV, accuracy is ±2%; Standard sensitivity: 10 mm/mV±0.2 mm/mV
Recording mode: auto mode, rhythm mode and manual mode
Measurement parameters: HR, P-R interval, P Duration, QRS Duration, T Duration, Q-T interval, Q-Tc, P Axis, QRS Axis, T Axis, R(V5), S(V1), R(V5)+S(V1)
Safety classification: Class I, type CF defibrillation-proof applied part
Polarization resistance voltage: ±610mV
Noise level: ≤12 µVp-p
ECG signal input sampling frequency: 32 kHz
Waveform data processing sampling frequency: 1 kHz
Sampling accuracy: 24-bit
The minimum detection signal: 10 Hz, 20 µV(peak-peak value) deflected sinusoidal signal can be detected
Pacing detection channel: standard II
Accuracy of input signal: ±5%
Amplitude quantization: ≤5µV/LSB
Interchannel time deviation: <100 µs
Power supply:
AC: 100V~240V(50/60Hz)
DC: 7.4V/2000mAh rechargeable lithium battery
Degree of protection against ingress of liquid: IPX0
Working mode: continuous working
Standard configuration:
A lead cable
A limb electrode
A chest electrode
A thermal recording paper
A power cord
An earth wire
A user manual
A power adapter
Physical characteristic
Working environment
Temperature: 5℃~40℃
Relative humidity: 25%~95%(non-condensing)
Atmospheric pressure: 700 hPa~1060 hPa
Transport and storage environment
Temperature: -20 ℃~+55 ℃
Relative humidity: ≤95%
Atmospheric pressure: 500 hPa~1060 hPa
Dimension: 207mm(L)×96mm(W)×62mm(H)
Weight: 0.5Kg